Quick Start RIGOL
User’s Guide for DSA1000 Series
1-9
Figure 1-5 Tracking generator output connection
CAUTION
Reverse power or voltage mustnt be higher than 1 W or 50 V DC to avoid
damages to the tracking generator.
3. PROBE POWER
This connector provides two lines of power supply to a RF probe: 15 V, 150 mA,
and -12.6 V, 150 mA.
4. RF INPUT 50Ω
The RF input may be connected to a device under measure via an N male
connector, see figure below.
Figure 1-6 RF input connection
CAUTION
To avoid damages to the instrument, signals at the input terminal must be
less than 50 V DC voltage components and 30 dBm (1 W) AC (RF)
components.