GD-70D - 106 -
Disconnect Gas
Disconnect the humidifier from the tubing
stub on the GAS IN fitting. Hold the humidifier
assembly by the clear tubes at the ends when
installing or removing the humidifier to avoid
damaging the humidifier. Do not hold the
humidifier tube itself.
Span Adjustment
Using the or keys, adjust the reading to
match the concentration shown in the “GD-
70D” column in the table at the beginning of
this section.
Press the SET key.
8.0ppm
SPAN VAL
MAINTENANCE
Span Adjustment Completed
The instrument will indicate whether the span
adjustment passed (OK) or failed (NG).
8.0ppm
SPAN OK
MAINTENANCE
OR
8.0ppm
SPAN NG
MAINTENANCE
Return to 2-2.SPAN
Save Adjustment
To save the adjustment result, press the SET
key.
(Press the MODE key to cancel this menu.)
8.0ppm
SPAN Y/N
MAINTENANCE
Adjustment Completed
After the span has been recorded, the
instrument will return to 2-2.SPAN
automatically.
8.0ppm
SPAN END
MAINTENANCE
Return to 2-2.SPAN
Returning to Detection Mode
Store the components of the calibration kit in a safe place.
Reconnect the dust filter and inlet line to the GAS IN fitting.
Press and hold the MODE key to return to Detection Mode.
CAUTION
Exhaust gas must be collected
in the exhaust bag or discharged
through the exhaust line.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GD-70D Series and is the answer not in the manual?

RKI Instruments GD-70D Series Specifications

General IconGeneral
BrandRKI Instruments
ModelGD-70D Series
CategoryGas Detectors
LanguageEnglish

Related product manuals