GD-70D - 106 -
Disconnect the humidifier from the tubing
stub on the GAS IN fitting. Hold the humidifier
assembly by the clear tubes at the ends when
installing or removing the humidifier to avoid
damaging the humidifier. Do not hold the
Using the ▲ or ▼ keys, adjust the reading to
match the concentration shown in the “GD-
70D” column in the table at the beginning of
this section.
SPAN VAL
Span Adjustment Completed
The instrument will indicate whether the span
adjustment passed (OK) or failed (NG).
SPAN OK
OR
SPAN NG
To save the adjustment result, press the SET
key.
(Press the MODE key to cancel this menu.)
SPAN Y/N
After the span has been recorded, the
instrument will return to 2-2.SPAN
automatically.
SPAN END
Returning to Detection Mode
Store the components of the calibration kit in a safe place.
Reconnect the dust filter and inlet line to the GAS IN fitting.
Press and hold the MODE key to return to Detection Mode.
CAUTION
Exhaust gas must be collected
in the exhaust bag or discharged
through the exhaust line.