- 73 - GD-70D
9-3 Fault Alarm Test "2-9" - "SET 4"
PW
A1
A2
F
LCD
From Detection Mode, press and hold
the MODE key for three seconds.
0.0ppm
SIH4
1-8.M MODE
Press the key until the Maintenance
Mode screen displays. (For GD-70D-EA
units, the Maintenance Mode screen is
item 1-9.)
1- 8
M MODE
MAINTENANCE
User Mode
In "1-8.M MODE", press the SET key.
1- 8
M MODE
MAINTENANCE
Then press and hold the SET key again
for three seconds.
- - - -
M MODE
MAINTENANCE
2-0.GAS TEST
The first item in Maintenance Mode will
display.
2- 0
GAS TEST
MAINTENANCE
2-9. SETTING1
Press the key until the SETTING1
screen displays.
Press the SET key.
2- 9
SETTING1
MAINTENANCE
SET 1.ALM P
Press the key until the F TEST
screen displays.
SET 1
F TEST
MAINTENANCE
SET 4.F TEST
Press the SET key.
SET 4
F TEST
MAINTENANCE