- 39 - GD-70D
<Detection Mode>
Flow Rate Indicator
Because the suction flow rate of the detector is automatically adjusted by the flow rate control function, the
flow rate, in principal, does not need to be controlled. As shown on the figure below, when the flow rate
does not correspond to the specified flow rate for some reasons, it is adjusted automatically.
5-5 Modes
Details on each mode are provided as follows. (* Operations are slightly different depending on the
detector type or sensor unit.)
Gas concentration
Gas name
- Gas concentration Perform the alarm test.
(see pg. 59)
Adjustment for oxygen)
(1-1 SPAN for
Perform the zero adjustment (span adjustment for oxygen).
Setting Display 1-2 CONFIRM Show the setting of typical menus.
•
First alarm setpoint (AL1)
•
Second alarm setpoint (AL2)
•
Alarm delay time
•
Zero suppression value
•
Zero follower ON/OFF
• Sensitivity correction ON/OFF
Show the current flow rate.
Show the program version of the main unit.
Show the program version of the sensor unit.
Show the program version of the communication function.
Maintenance Mode
1-8 M MODE Enter maintenance mode.
WARNING
• When the detector enters each mode from Detection Mode while an alarm is activated, the alarm
contact is released.
CAUTION
• If the automatic flow rate adjustment does not work (due to clogged tube or leak), you may get a
"FLOW" message (for an unstable flow rate) or an "E-05" message (for flow rate abnormalities). In
this case, you must identify the causes and take appropriate actions.
too low
too high