6-4 Other Functions
<Suppression Function>
Some sensors are influenced by environmental changes (temperature, humidity, and other characteristics)
or interference gases which may cause the reading to vary in fresh air.
The suppression function is used to hide (suppress) the variation of the reading under the set suppression
value, so that the display indicates zero (or 20.9% for oxygen) until the reading exceeds the set
suppression value.
The factory setting for the suppression function is OFF for every gas type except for oxygen.
<<Example: SiH4 0 - 15 ppm range>>
0.5ppm
SIH4
Suppression Disabled
A variation around zero is displayed as the reading.
0.0ppm
SIH4
Suppression Enabled
A variation under the set suppression value (ie. 6% of full scale or 0.9 ppm for
SiH4) is shown as zero.
CAUTION
If the zero reading is BELOW zero, the reading will not change until the reading reaches -10% of full
scale.
A reading that is -10% of full scale is displayed as "-0.0".

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GD-70D Series and is the answer not in the manual?

RKI Instruments GD-70D Series Specifications

General IconGeneral
BrandRKI Instruments
ModelGD-70D Series
CategoryGas Detectors
LanguageEnglish

Related product manuals