10-8 Flow Rate Adjustments
<Flow Rate Manual Adjustment "2-6">
The 2-6 FLOW menu item allows the user to manually increase or decrease the flow. The flow rate of the
detector is normally automatically adjusted to 0.5 L/min. In order to perform a manual adjustment, you
must first turn off the auto-adjustment function (found in Maintenance Mode item 2-10 SET-11).
From Detection Mode, press and hold the
MODE key for three seconds.
SIH4
1-8.M MODE
Press the ▲ key until the Maintenance
Mode screen displays.
M MODE
In "1-8.M MODE", press the SET key.
M MODE
Then press and hold the SET key again for
three seconds.
M MODE
The first item in Maintenance Mode will
display.
GAS TEST
2-6. FLOW
Press the ▲ key until the FLOW screen
displays.
FLOW
MAINTENANCE
Use the ▲ or ▼ key to adjust the flow rate
to 0.5 L/min.
Press the SET key.
(Press the MODE key to cancel this menu.)
0.50 L/M