GD-70D - 110 -
10-8 Flow Rate Adjustments
<Flow Rate Manual Adjustment "2-6">
The 2-6 FLOW menu item allows the user to manually increase or decrease the flow. The flow rate of the
detector is normally automatically adjusted to 0.5 L/min. In order to perform a manual adjustment, you
must first turn off the auto-adjustment function (found in Maintenance Mode item 2-10 SET-11).
PW
A1
A2
F
LCD
From Detection Mode, press and hold the
MODE key for three seconds.
0.0ppm
SIH4
1-8.M MODE
Press the key until the Maintenance
Mode screen displays.
1- 8
M MODE
MAINTENANCE
User Mode
In "1-8.M MODE", press the SET key.
1- 8
M MODE
MAINTENANCE
Then press and hold the SET key again for
three seconds.
- - - -
M MODE
MAINTENANCE
2-0.GAS TEST
The first item in Maintenance Mode will
display.
2- 0
GAS TEST
MAINTENANCE
2-6. FLOW
Press the key until the FLOW screen
displays.
Press the SET key.
2- 6
FLOW
MAINTENANCE
Flow Rate Adjustment
Use the or key to adjust the flow rate
to 0.5 L/min.
Press the SET key.
(Press the MODE key to cancel this menu.)
50 %
0.50 L/M
MAINTENANCE