- 109 - GD-70D
CAUTION
Do not press the SET key before the reading is stabilized.
Press the SET key when the reading has
stabilized.
* The reading must be above 10% of the full
scale value in order to continue.
SPAN GAS
MAINTENANCE
Disconnect the calibration tubing from the
tubing stub on the pyrolyzer’s GAS IN fitting.
Using the ▲ or ▼ keys, adjust the reading to
match the concentration shown in the “GD-
70D” column in the table at the beginning of
this section.
SPAN VAL
Span Adjustment Completed
The instrument will indicate whether the span
adjustment passed (OK) or failed (NG).
SPAN OK
OR
SPAN NG
To save the adjustment result, press the SET
key.
(Press the MODE key to cancel this menu.)
SPAN Y/N
After the span has been recorded, the
instrument will return to 2-2.SPAN
automatically.
SPAN END
Returning to Detection Mode
Store the components of the calibration kit in a safe place.
Reconnect the dust filter and inlet line to the GAS IN fitting on the pyrolyzer.
Press and hold the MODE key to return to Detection Mode.
CAUTION
Exhaust gas must be collected
in the exhaust bag or discharged
through the exhaust line.