- 111 - GD-70D
<Flow Rate Default Set "2-5">
Regardless of whether the flow rate is manually or automatically set, when the reading on the flow rate
indicator is incorrect (due to aging degradation, improper flow rate default set, or other reasons), it must be
set to provide the right flow rate.
To set the Flow Rate Indicator, prepare and connect a flow meter (which indicates 0.5 L/min precisely),
and perform the default set with the flow meter while it is indicating 0.5 L/min.
From Detection Mode, press and hold the
MODE key for three seconds.
SIH4
1-8.M MODE
Press the ▲ key until the Maintenance
Mode screen displays.
M MODE
In "1-8.M MODE", press the SET key.
M MODE
Then press and hold the SET key again for
three seconds.
M MODE
The first item in Maintenance Mode will
display.
GAS TEST
2-5. SPAN
Press the ▲ key until the DEF FLOW
screen displays.
DEF FLOW
MAINTENANCE
Flow Rate Adjustment/Default Set
Adjust the flow rate by pressing the ▲ or ▼
key until the connected flow meter indicates
0.5 L/min ± 10%.
Press the SET key. (MODE to cancel)
DEF FLOW
CAUTION
Perform the flow rate default set after verifying that the flow rate is 0.5 L/min by using a flow meter.