- 111 - GD-70D
<Flow Rate Default Set "2-5">
Regardless of whether the flow rate is manually or automatically set, when the reading on the flow rate
indicator is incorrect (due to aging degradation, improper flow rate default set, or other reasons), it must be
set to provide the right flow rate.
To set the Flow Rate Indicator, prepare and connect a flow meter (which indicates 0.5 L/min precisely),
and perform the default set with the flow meter while it is indicating 0.5 L/min.
PW
A1
A2
F
LCD
From Detection Mode, press and hold the
MODE key for three seconds.
0.0ppm
SIH4
1-8.M MODE
Press the key until the Maintenance
Mode screen displays.
1- 8
M MODE
MAINTENANCE
User Mode
In "1-8.M MODE", press the SET key.
1- 8
M MODE
MAINTENANCE
Then press and hold the SET key again for
three seconds.
- - - -
M MODE
MAINTENANCE
2-0.GAS TEST
The first item in Maintenance Mode will
display.
2- 0
GAS TEST
MAINTENANCE
2-5. SPAN
Press the key until the DEF FLOW
screen displays.
Press the SET key.
2- 5
DEF FLOW
MAINTENANCE
Flow Rate Adjustment/Default Set
Adjust the flow rate by pressing the or
key until the connected flow meter indicates
0.5 L/min ± 10%.
Press the SET key. (MODE to cancel)
1000
DEF FLOW
MAINTENANCE
CAUTION