- 37 - GD-70D
<Warm-up Sequence>
The warm-up taking approximately 25 seconds to complete.
Power On
PW
A1
A2
F
LCD
Initial Clear
──────
WARM UP
Gas
Specifications
Display
15.0ppm
SIH4
Detection Mode
0.0ppm
SIH4
WARNING
When Oxygen (OSU) is selected, fresh air (20.9 vol%) is
the output after the detector is started in the atmosphere.
For a specification where an alarm is triggered by the
AIR (normal oxygen concentration) output (e.g., 0 - 5
vol%), be careful of an alarm issue after the detector is
started.
CAUTION
Do not turn off the detector during the initial clear. The detector is reading the sensor memory during
the initial clear.
If a new sensor unit is installed or the sensor unit is replaced after the detector is started, the sensor
unit must be warmed up for a specified period which is determined depending on the type of the
sensor unit. Suggested warm-up time for SGU type sensor units is displayed in Maintenance Mode.
During the warm-up, the alarm activation and output signals are unstable. Prepare appropriately for
false alarms.
Because the pyrolyzer unit (PLU-70) must be warmed up for one hour, please warm it up with the
detector.
After the warm-up is completed, verify that the reading on the flow rate indicator corresponds to the
specified flow rate, and then perform a gas calibration.
PW:POWER
A1:ALM1
A2:ALM2
F:FAULT
:Lamp on
:Lamp off

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GD-70D Series and is the answer not in the manual?

RKI Instruments GD-70D Series Specifications

General IconGeneral
BrandRKI Instruments
ModelGD-70D Series
CategoryGas Detectors
LanguageEnglish

Related product manuals