R&S FSH3 Setting and Using the Measurement Functions
The R&S FSH3 then requests to terminate the output
of the tracking generators into 50 Ω.
¾ Connect the generator output with the
50 Ω termination.
¾ Press the F1 or F2 (CONTINUE) softkey.
During calibration, the R&S FSH3 outputs the
message "Calibrating LOAD, please wait".
When calibration is over, Transmission (vect) is
displayed in the top right-hand corner to indicate that
the R&S FSH3 is vector-calibrated for transfer
measurement.
Connect the DUT between the tracking generator
output and the RF input. The R&S FSH3 displays the
magnitude of the transfer function of the DUT.
1145.5973.12 4.59 E-6