EasyManua.ls Logo

Rohde & Schwarz FSH3 - Reflection Measurements

Rohde & Schwarz FSH3
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setting and Using the Measurement Functions R&S FSH3
Reflection measurements
The VSWR Bridge R&S FSH-Z2 should be used for reflection measurements. However, an equivalent
bridge (e.g. the ZRB2 from Rohde & Schwarz) can be used. The R&S FSH-Z2 bridge is directly
screwed to the RF input connector and the generator’s output.
¾ Connect the control cable of the R&S FSH-Z2 with the socket Power Sensor of the R&S FSH3.
¾ Connect the RF and Generator port of the R&S FSH-Z2 to the RF input and generator output of the
R&S FSH3.
Test port
Control
cable
VSWR Bridge and
Power Divider
R&S FSH-Z2
The R&S FSH3 can measure both a DUT’s return loss and its VSWR.
Scalar measurement of reflection
The test setup must be calibrated before any measurements are made. This is done with a short and an
open at the point were the reflection measurement is to be made. If a cable is to be inserted between
the DUT and the bridge, perform the calibration at the measurement end of the cable.
¾ Press the REFLECT CAL softkey.
The R&S FSH3 prompts the user to connect an open
to the measurement input.
¾ Connect an open to the measurement port of the
bridge.
¾ By pressing the first or the second softkey
(CONTINUE), start the OPEN calibration. While
calibration is in progress, the R&S FSH3 outputs
the message "Calibrating OPEN, please wait... ".
¾ Press the CANCEL softkey to abort calibration.