Using the Markers R&S FSH3
Operating sequence:
¾ Press the MARKER key.
¾ Press the SET MARKER softkey.
¾ Using the cursor keys or the rotary knob, select the
function you want.
¾ Confirm your selection with the ENTER key or the
SET MARKER softkey.
The R&S FSH3 then performs the action you have
selected.
Using more than one marker at a time (multimarker mode)
To measure different signals in a trace, the R&S FSH3 has the multimarker function. Up to six different
markers are available in the multimarker mode. Marker 1 measures in absolute units. Markers 2 to 6
can measure in absolute units (marker) as well as relative units (delta). The reference for delta markers
is always marker 1.
Operating sequence:
¾ Press the MARKER key.
¾ Press the MARKER MODE softkey.
¾ Using the cursor keys or the rotary knob, select
MULTI MARKER... from the submenu.
¾ Confirm your selection with the ENTER key or the
MULTI MARKER softkey.
The R&S FSH3 is now in multimarker mode.
Except for the MARKER DEMOD softkey that is
replaced by the MKR LIST VIEW softkey in
multimarker mode, this menu is identical to the
standard marker menu. The marker designation
contains the number of the marker in question (M
becomes M1, D becomes D2). The active marker or
delta marker is displayed with its number (e.g. M1: or
D2:) and the frequency and level at the upper-right
hand corner of the screen.