R&S FSH3 Saving and Recalling Settings and Test Results
Saving Calibration Data
When performing scalar transmission or reflection measurements, the R&S FSH3 can store the
calibration data along with the settings and results. Saving the settings and results with calibration data
requires twice as much memory space as without it. This, of course, reduces the maximum number of
data sets that can be saved.
In the default state, calibration data storage is disabled.
¾ Press the SETUP key.
¾ Press the GENERAL softkey.
¾ Select SAVE CAL DATA... and confirm your
selection by pressing the ENTER key or the
GENERAL softkey.
¾ Using the rotary knob or the cursor keys, select ON
or OFF.
¾ Confirm with ENTER.
The state of calibration data saving is entered in the SETUP menu.
When recalling data sets with calibration data included the R&S FSH3 checks, if the current instrument
temperature corresponds to the instrument temperature at the time of calibration. If it deviates more
than 5 °C the R&S FSH3 outputs a red dot in front of the • Transmission or • Reflection display. A
new calibration is recommended, then.
1145.5973.12 2.21 E-6