Switchover External Reference / External Trigger R&S FSH3
¾ Press the GENERAL key.
The R&S FSH3 opens the submenu with the general
settings. The cursor is on the POWER DOWN menu
item.
¾ Confirm the POWER DOWN selection by pressing
the ENTER key.
The R&S FSH3 opens a selection window with the
settings: 5 minutes, 30 minutes and DISABLE.
¾ Use the Rotary knob or the Cursor keys to select the
setting you want and confirm by pressing the
ENTER key or the GENERAL softkey.
Switchover External Reference / External Trigger
The Ext Trig/Ext Ref BNC connector on top of the R&S FSH3 can be used either as an input for an
external trigger or an external reference. Switchover is via the Setup menu.
¾ Press the SETUP key.
¾ Press the HARDWARE SETUP softkey.
The active input setting is highlighted in green.
¾ Use the Rotary knob or the Cursor keys to select
EXT REF or EXT TRIG.
¾ Confirm the selection with the ENTER key or the
HARDWARE SETUP softkey.
The EXT TRIG setting is only for input configuration. The use of the external trigger must be set in the
Sweep menu (SWEEP key, TRIGGER softkey).
The input setting can be queried via the status display (press STATUS key).