Setting and Using the Measurement Functions R&S FSH3
When REFLECTION is selected, the R&S FSH3
measures the return loss over the frequency range
which has been selected for the distance-to-fault cable
measurement. This means, for example, an antenna
can be matched without altering the test setup. The
R&S FSH3 automatically switches the VSWR Bridge
R&S FSH-Z2 over to the VSWR measurement mode if
REFLECTION has been selected.
To indicate that the R&S FSH3 is measuring return
loss, DTF refl. cal is displayed in the top right-hand
corner of the screen.