Measuring with Transducer Factors R&S FSH3
If a transducer with the unit dBµA/m is switched on, no other unit can be selected in the AMPT menu.
Level indication is entirely in dBµA/m.
Reference level settings for measurements with transducers:
The transducer shifts the trace by its value as a function of frequency. Positive transducer values
increase the level, negative values reduce it. To ensure that the trace is always within the diagram, the
R&S FSH3 adjusts the reference level accordingly. The reference level is shifted by the maximum
transducer value in the positive or negative direction.
Frequency range of transducer:
If the set frequency range is wider than the span in which a transducer is defined, the R&S FSH3
assumes the transducer values outside the defined range to be zero.
Data sets containing transducer factors:
The R&S FSH3 stores data sets together with any transducer factors that may have been active for the
measurement in question. When such a data set is recalled, the associated transducer factor(s) are
switched on as well. Transducer factors recalled as part of a data set do however not appear in the list
of transducer factors.