R&S FSH3 Measurements on CW Signals
When saving a data set for a scalar reflection measurement in a calibrated state, the R&S FSH3 can
store the calibration data along with the other settings (see section "Saving Calibration Data"). Thus
after recalling the setting, a measurement can be performed without prior calibration, providing the
instrument's temperature does not deviate more than 5 °C from its temperature when the data set was
stored.
If the temperature deviation is greater, the R&S FSH3 outputs a (red) dot in front of Reflection . A
precise measurement can then only be made after a calibration.
1145.5973.12 2.13 E-6