Remote Commands
R&S
®
FSW
807User Manual 1173.9411.02 ─ 19
INPut:LISN:TYPE <LISNType>
This command selects the type of network.
Parameters:
<LISNType> TWOPhase
2-Phase networks
FOURphase
4-Phase networks
ESH3z5
ESH2z5
ENV4200
ENV216
*RST: ---
Example:
INP:LISN:TYPE ESH3z5
Selects the ESH3z5 network.
Manual operation: See "LISN Type" on page 305
11.5.13.6 Retrieving EMI Results
The commands required to retrieve EMI measurement results are described here.
Useful commands for retrieving EMI measurement results described elsewhere:
● CALCulate<n>:MARKer<m>:X on page 932
CALCulate<n>:DELTamarker<m>:X on page 931
● CALCulate<n>:MARKer<m>:Y? on page 943
CALCulate<n>:DELTamarker<m>:Y? on page 943
Remote commands exclusive to retrieving EMI measurement results
CALCulate:DELTamarker<m>:FUNCtion:FMEasurement:RESult?...................................... 807
CALCulate:MARKer<m>:FUNCtion:FMEasurement:RESult?..............................................808
CALCulate:DELTamarker<m>:FUNCtion:FMEasurement:LIMit<k>:CONDition ?.................. 808
CALCulate:MARKer<m>:FUNCtion:FMEasurement:LIMit<k>:CONDition ?.......................... 808
CALCulate:DELTamarker<m>:FUNCtion:FMEasurement:LIMit<k>:DELTa?.........................809
CALCulate:MARKer<m>:FUNCtion:FMEasurement:LIMit<k>:DELTa?.................................809
CALCulate:DELTamarker<m>:FUNCtion:FMEasurement:RESult? <Result>
This command queries the result of the EMI measurement at the deltamarker position.
Return values:
<Result> Absolute power level measured during final EMI measurement.
Example:
CALC:DELT3:FUNC:FME:RES?
Queries the result of delta marker 3.
Usage: Query only
Configuring and Performing Measurements