Ultrafilter Installation
Caution!
Danger of irreversibly damaging
arium 611VF components and the
filter cartridges! Tighten all union
nuts made of plastic by hand. 
Do not use a wrench or pliers.
Before starting up arium 611VF,
you need to install the ultrafilter
module in the housing cover. 
Proceed as follows:
• Make sure to unplug the power
cord from the electrical outlet (“O”).
• Open both doors of the unit.
• Remove ultrafilter from packaging
and remove protective covers from
fittings.
• Connect the tubing connectors to
the ultrafilter module as shown in
the Figure:
(1): green connector for feed 
water - Bottom
(2): blue connector for product
water - Upper left
(3): red connector for reject 
water - Top
• Push in the ultrafilter module all
the way into the holders so that 
the connector fitting on the side 
is facing left.
• Close both doors.
10
!!
Attach the three tubing connectors to the ultrafilter module 
1
2
3