32
Specifications
Dimensions W + D + H 43.2 + 48.3 + 34.3 [cm]
Empty weight 15.9 kg
Operating weight 23.6 kg
Clearance requirements Sides: 10 cm minimum for cable and tubing connections
Front: 40 cm minimum for opening the doors
Inlet water requirements arium 611VF requires water pretreated by:
- distillation
- deionization or
- reverse osmosis
Distilled water: > 250 KO · cm (< 4 µS/cm)
RO water: TDS < 25 ppm CaCO
3
> 20 KO · cm (< 50 µS/cm)
Deionized water: TDS < 10 ppm CaCO
3
> 50 KO · cm (<20 µS/cm),
All: Turbidity < 1 N.T.U.
Silica < 1000 ppb
TOC < 1000 ppb
Pressure: From atmospheric tank inlet to a maximum inlet pressure of 7 bar
Product water capabilities Quality of the
purified water: 18.2 MO + cm at 25°C
RNase-/DNase-/DNA-free
Bacteria: < 1 CFU/100 ml
TOC: < 1 ppb
Endotoxin: < 0,001 EU/ml
Flow rate Up to 1.5 l/min maximum at minimum inlet feed water pressure of 2.0 bar at 50 HZ and
with a new final filter.
Environmental conditions Operating: 5°C - 28°C; 80% relative humidity, non-condensing
Storage: 5°C - 45°C; 80% relative humidity, non condensing
Electrical requirements 100-240 V~, 50/60 Hz, 1 phase
Fuses power entry 2 fuses 5 + 20 mm, time lag, 250 V, 1 A
Installation Category II (over-voltage) in accordance with IEC 664,
Pollution Degree 2 in accordance with IEC 664.
Altitude limit: 5,000 meters.
Complies with the EMV: EN 50081-1, EN 50082-1
following standards Safety: IEC 1010-1-92