Remove screw from metal shield
Remove UV bulb (4) from stainless steel chamber
29
Activate the UV Timer
The UV bulb should be replaced
approximately every six months.
A timer can be activated to display
a message every 6 months to
remind you that the UV bulb is
ready to be replaced.
To activate the timer after installing
a new UV bulb:
• Press MENU to get into the Main
Menu:
• Select “Maintenance” and press
ENTER to confirm.
• Select “UV Timer” by using y
and press ENTER to confirm.
• Press ENTER to reset the timer.
The internal timer is reset and will
remind you to replace the UV bulb
after 6 months.
• Alternatively press the MENU key
to return to the Main Menu without
resetting the timer.
After 6 months the display will
show the following:
18.2 MO * cm c
Change UV Bulb
09.48 23.04.01
UV LIFE TIMER
Timer Reset
Escape
MAINTENANCE
UV Life Timer
Sanitization
MENU Utilities
Maintenance
Setup
1234 5