EasyManua.ls Logo

Sartorius arium 611VF - Activate the UV Timer

Sartorius arium 611VF
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remove screw from metal shield
Remove UV bulb (4) from stainless steel chamber
29
Activate the UV Timer
The UV bulb should be replaced
approximately every six months.
A timer can be activated to display
a message every 6 months to
remind you that the UV bulb is
ready to be replaced.
To activate the timer after installing
a new UV bulb:
Press MENU to get into the Main
Menu:
Select “Maintenance” and press
ENTER to confirm.
Select “UV Timer” by using y
and press ENTER to confirm.
Press ENTER to reset the timer.
The internal timer is reset and will
remind you to replace the UV bulb
after 6 months.
Alternatively press the MENU key
to return to the Main Menu without
resetting the timer.
After 6 months the display will
show the following:
18.2 MO * cm c
Change UV Bulb
09.48 23.04.01
UV LIFE TIMER
Timer Reset
Escape
MAINTENANCE
UV Life Timer
Sanitization
MENU Utilities
Maintenance
Setup
1234 5