Technical Data
15.7.5 Models MCA623S | MCA623P | MCA323S | MCA5202S
MCA623S MCA623P MCA323S MCA5202S
Unit Value Value Value Value
Scale interval (d) mg 1 1 | 2 | 5 1 10
Maximum capacity (Max) g 620 150 | 300 |
620
320 5200
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg 0.7 1 0.7 6
Standard deviation of the load values,
typical value
mg 0.4 0.4 0.4 2
Repeatability at approx. maximum capacity
Standard deviation of the load values,
tolerance
mg 0.7 1 0.7 6
Standard deviation of the load values,
typical value
mg 0.5 0.5 0.5 2
Linearity deviation
Tolerance mg 2 5 2 10
Typical value mg 0.6 1.5 0.6 5
Deviation when load is off-center,
positions according to OIML R76
Test weight g 200 200 200 2000
Tolerance mg 2 4 2 10
Typical value mg 1 3 1 5
Sensitivity drift between +10°C and +30°C ppK 2 2 2 2
Tare maximum capacity: Less than 100%
ofmaximum capacity
Accuracy class according
toDirective2014/31/EU
II II II I
Verification scale interval (e) according
toDirective 2014/31/EU
mg 10 10 10 100
Minimum load (Min) according
toDirective2014/31/EU
mg 20 20 20 1000
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 820 820 820 8200
Typical minimum initial weighing mg 820 820 820 8200
Typical stabilization time s 0.8 0.8 0.8 0.8
Typical measurement time s 1 1 1 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA Series and is the answer not in the manual?

Sartorius Cubis MCA Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA Series
CategoryScales
LanguageEnglish

Related product manuals