90 Operating Instructions Cubis
®
MCA
Operation
8.10 Adjusting with the isoCAL Function
The device can be automatically internally calibrated and adjusted using
theisoCAL function.
Requirements
The device is not located in the menu.
Alphanumeric inputs are not active.
The load on the weighing pan remains unchanged for 2 minutes.
The load on the weighing pan amounts to no more than 2% of the
maximum load.
The device does not register an input for 2 minutes.
If all requirements for starting the isoCAL function are met and one of the
following conditions occurs: The isoCAL function is automatically triggered.
Possible conditions are:
The ambient temperature has changed since the last adjustment.
The interval time was exceeded (interval time, see Chapter “15.9 isoCAL
Function”, page 116).
The device has been leveled.
The device has been disconnected from the power supply since the last
adjustment (only for conformity-assessed models).
Procedure
t If the isoCAL automatic start function is set and the [isoCAL] button
isdisplayed as the predominant button in the operating display:
t Wait until the isoCAL function starts automatically.
y In the operating display, a time display counts down to 0.
y If no load change or no operation takes place on the device before
the expiration of the time display: The isoCAL function starts.
t If the isoCAL manual start function is set and the [isoCAL] button
isdisplayed as the predominant button in the operating display:
t Press the [isoCAL] button.
t If information is displayed stating that the weighing pan is loaded:
Unload the weighing pan.
y The isoCAL function starts.
y If the isoCAL function is complete: The device confirms the completion
of the calibration | adjustment process with an acoustic signal,
andthecalibration report is displayed.
t Output the calibration report via a connector: Press the [Print Memory]
button.
t To close the calibration report and return to the previous display:
Pressthe [OK] button.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA Series and is the answer not in the manual?

Sartorius Cubis MCA Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA Series
CategoryScales
LanguageEnglish

Related product manuals