Operating Instructions Cubis
®
MCA 71
System Settings
7.7.9 Parameters in the “Connections / Website / Web Services” Menu
Parameter Setting Values Explanation
Website access Deactivated Deactivates the display of the website for the device.
On, without
authentication*
Sets the display options of the website for the device
to“withoutauthentication”.
Remote access View only* Permits remote access in view mode via a browser.
View and remote control Permits remote access via a browser.
Web service
password
User input Saves a password for web service access.
* Factory setting
7.7.10 Parameters in the “Connections / Certificates” Menu
Parameter Setting Values Explanation
Trusted
certification
authorities
List of certification
authorities
Activates a trusted certification authority (CA), in order to
identify trusted servers, e.g., “GlobalSign”.
Loading certification
authorities from
aconnector
Imports a certification authority, e.g., using a USB stick.
Device certificates List of device certificates Activates a device certificate, e.g “GlobalSign”.
Loading device
certificates from
aconnector
Imports a certification authority for safe remote access
tothedevice website, e.g., from a network server or USB stick.
Trust unknown
certification
authorities
No (secure)
Yes (insecure)*
* Factory setting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA Series and is the answer not in the manual?

Sartorius Cubis MCA Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA Series
CategoryScales
LanguageEnglish

Related product manuals