94 Operating Instructions Cubis
®
MCA
Operation
8.14 Turning the Ionizer On | Off
(OnlyforDevices with an Ionizer)
8.14.1 Setting the Ionizer
Procedure
t Open the “Settings / Device Settings / Ionizer” menu.
t Select manual or automatic activation for the “Ionizer Function”
parameter.
t For the “Ionizer Intensity” parameter, select the desired intensity,
e.g.weak.
t For the “Operating Duration” parameter, select the duration of the
ionization process, e.g. 60 seconds.
8.14.2 Starting the Ionization Process
Requirements
The model is equipped with an ionizer.
Procedure
t If the [Ionizer] button appears in the weighing display:
Pressthe[Ionizer]button.
y The ionization process starts.
8.14.3 Switching Off the Ionizer
Procedure
t Open the “Settings / Device Settings / Ionizer” menu.
t For the “Ionizer Function” parameter, select the setting value “Off,
nofunction”.
8.15 Running Applications (Examples)
8.15.1 Executing the “Unit Change” Function
The “Unit Change” function enables the user to switch between
thedifferent units and resolutions defined in the weighing profile of the
activetask. The units and resolutions can be set at the beginning of the
weighingprocess.
Procedure
t Start the desired task.
t Press the [Unit Change] button.
y All units defined in the weighing profile for the active task are displayed
in a list.