Operating Instructions Cubis
®
MCA 91
Operation
8.11 Internally Calibrating and Adjusting
theDevice
Requirements
The weighing pan is unloaded.
Procedure
t Open the main menu.
t Select the “Internal Adjustment” task.
y The internal calibration | adjustment function is executed.
y If automatic leveling is set for a model with motorized adjusting feet:
Thedevice levels itself automatically.
y If the calibration | adjustment function is complete: The device confirms
the completion of the calibration | adjustment process with an acoustic
signal, and the calibration report is displayed.
t Output the calibration report via a connector: Press the [Print Memory]
button.
t To close the calibration report and return to the main menu:
Pressthe[OK] button.
8.12 Weighing and Printing with ID Marking
8.12.1 Saving Values for Printouts
Every lot and every sample can be assigned an ID number. The ID numbers
are saved in the print memory and exported during the printing process.
Requirements
The prompt for the lot IDs and sample IDs is activated for the
“Standardweighing” task.
Procedure
t Open the main menu.
t Start the “Weighing” task: To do so, tap on the task.
y The input field for the lot ID is displayed.
t If a barcode scanner is connected to the device: Scan or type the lot ID
into the entry field.
t If no barcode scanner is connected to the device: Type the lot ID into
theentry field.
t Press the [OK] button.
y The weighing display is displayed.
t Zero the device. In order to do so, press the [Zero] button.
t Place the sample on the weighing pan.
t Press the [Confirm] button.
y The input field for the sample ID is displayed.
t Scan or type the lot ID into the entry field.
t Press the [OK] button.
y The [Print memory] button appears in the operating display.
y The weight value and entered IDs are saved.