TARE FUNCTION
Starting point for a tare function is the operating level:
+
Prerequisite at configuration level 3
Corr = 1 r page 9
TARE:
Watch display (container tare weight signal) and wait, until the measurement
value is stable.
Press key D 5 during 5 seconds.
Subsequently, the following value is displayed : -0-
This operation can be repeated several times.
Deleting:
Press key D during approx. 15 seconds.(After 5 seconds, the tare value is
displayed:
-0-.) Only after this tare display has stopped, the tare correction
was deleted and key D can be released.
g
After changing the sensor, the decimal point or the measuring range (change in
Con1), another taring operation must be carried out, or the current one must be
deleted.
+
After measurement value correction, the access should be disabled again.
Corr = 0 r page 9
Another simple possibility for tare adjustment is by connecting terminals 2 and 3
shortly.
A blocking function for this can be realized by means of key switch.
+
9408-800-25001 15