4� At the “Add sample to pan” prompt, open the heater
hood and begin adding sample to the pan� A bar graph
on the display will guide the operator to place the
correct amount of sample as set by the ideal weight
parameter of the drying program� Continue adding
sample until the beeper sounds, indicating that the
sample amount is within 10% of the ideal weight� The
exact weight is also displayed on the screen�
5� At the “Close heater hood” prompt, close the heater
hood� The balance will capture the initial weight of the
sample “Sampling” and then show that the test is in
progress� The display will change to show
current information including the elapsed time and the
current percentage weight loss in the units as specified
in the drying program�
Note
To stop the test in progress, press the Start|Stop key�
The printer will begin to print as the test begins� Print-
out information can be selected in Setup menu under
Printer (LMA100P only)�
At the end of the test the display will show “Test
Done�” The result will stay on the screen as well as be
printed� The analyzer will enter the Standby mode until
the next test is initiated�
27
Dry Milk Powder 09/07/2012 02:14 PM
Test in Progre ss
0.015
% Moisture
Elapsed time: 00.10
Graph
Press STOP key to stop test
Temp 1
Current
105
71
Press START ke y to begin test
Dry Milk Powder 09/07/2012 02:14 PM
Test Done
3.471 % Moisture
Result time: 7.0 min
ID: 126399.1
Target: Pass
Recall Setup
Graph User
Wa rning: Sample on pa n
Standby
Current
105
105

Table of Contents