EasyManua.ls Logo

Sartorius LMA100P - Mark 3 Program Download Format

Sartorius LMA100P
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mark 3 Program Download Format
Introduction: This document details the program output format for the Mark 3 Moisture
Analyzer� The analyzer has the capability to transfer program settings to another Mark 3
or a PC� The analyzer can also be downloaded with settings from a PC�
This text format is used to transfer all the information necessary to control the test
conditions of a moisture analysis test�
Field Description
1 Program Number (#XXX)
2 Program Name (15 characters max�)
3 Units (0=Moisture, 1=Solids, 2=Volatiles, 3=Moisture dry, 4=PPM Moisture,
LMA100P: 5=PPM Solids, 6=Weight, 7=mg/L
LMA110S: 5=Weight, 6=mg/L
4 Slope Time (X�X minutes)
5 Slope Percentage (X�XXX)
6 Slope Calculation (0=Actual, 1=Calc 1, 2=Calc 2, 3=Calc 3)
7 Standby Temperature (165C max)
8 Equilibration Time (X minutes)
9 Ideal Weight (XX�X grams)
10 Ideal Weight % (XX�X)
11 Ideal Weight Lock (0=OFF, 1=ON)
12 Target Limit: Upper (X�XXX %)
13 Target Limit: Lower (X�XXX %)
14 Target Limit: Usage (0=OFF, 1=ON)
15 Test Resolution (2=X�XX%, 3=X�XXX%)
16 Scale Factor (X�XXX) Default=1�000
17 Scale Offset (X�XXX) Default=0�000
18 Temperature 1 (XXX)
19 Ramp 1 (0=Standard, 1=Rapid)
20 Time 1 (XX�X)
21 Temperature 2 (XXX)
22 Ramp 2 (0=Standard, 1=Rapid)
23 Time 2 (XX�X)
24 Test Mode (0=Standard, LMA100P: 1=Syringe, 2=Concentration, 3=Fat
LMA110S: 1=Concentration, 2=Fat)
25 Start Delay (seconds)
Output Example:
# 1,Default,0,1�0,0�050,0,60,0,5�0,10�0,0,100�00,0�000,0,3,1�000,0�000,105,0,0�0,0,0,
0�00,128/r/n
59