Mark 2 Program Download Format
Field Description
1 Program Number (#XXX)
2 Units (M,S,W,V,L)
3 Ideal Weight (XX�X, Negative value indicates ideal wt� loading locked 90-110%)
4 Temp 1 (XXX)
5 Temp 2 (XXX)
6 Time 1 (Seconds)
7 Time 2 (Seconds)
8 Standby Temp (XXX)
9 Slope Mode (0=Actual, 1=Calc1, 2=Calc2, 3=Calc3)
10 Slope Time (Tenths of minutes, example: 50=5�0 minutes)
11 Slope % (XX�XXX)
12 Start Delay of Initial Wt� Capture*
13 Program Name (19 characters)
Output Example (Factory Defaults):
# 1,M, 5�0,105, 0, 0, 0, 60,0,10, 0�050,65,FACTORY 1
# 2,M,30�0,110, 0, 120, 0,110,0,20, 0�010,65,FACTORY 2
# 3,S, 2�0,135, 0, 0, 0,130,0,10, 0�100,65,FACTORY 3
# 4,W, 5�0,105,130, 300, 300, 60,0, 0, 0�000,65,FACTORY 4
# 5,L, 2�0,135, 0, 0, 0,100,0,10, 0�050,65,FACTORY 5
/r/n
* Field 12 - Start Delay of Initial Wt� Capture Bit 0���Sample ID 0=Off 1=On
Sample ID Off Sample ID On
Time(s) Value Hex Time (s) Value Hex
0 0 0000 0000 0 1 0000 0001
1 16 0001 0000 1 17 0001 0001
2 32 0010 0000 2 33 0010 0001
3 48 0011 0000 3 49 0011 0001
4 64 0100 0000 4 65 0100 0001
5 80 0101 0000 5 81 0101 0001
6 96 0110 0000 6 97 0110 0001
7 112 0111 0000 7 113 0111 0001
8 128 1000 0000 8 129 1000 0001
9 144 1001 0000 9 145 1001 0001
60