76
§ The automatic temperature adjustment report is printed
after the adjustment process�
To verify an adjustment temperature:
§ From the Service menu, press the 3 key to display the
Adjust Temperature screen�
§ Press the 2 for Verify temperature� The screen will
change to the Verify temperature screen�
§ The default verifies temperature is 130°C� Press the
“Temp” soft key to begin� The screen will change to the
Temp verify in process screen and show the selected
temperature� After 20 minutes, the screen will change
to the Temp verify done screen and will display the
expected selected temperature and the actual tempera-
ture that was measured by the
temperature bowl (“fixture”)�
§ Press the Print soft key to print a temperature
verification report (LMA100P only)�
§ If the agreement of the selected temperature and the
actual temperature are not within 2 degrees for any of
the selected temperatures, it is recommended that you
perform the temperature calibration and adjustment
procedure�
Automatic Temperature Adjustment Report
- - - - - - - - - - - - - - - - - - - - - - - - - -
Fixture number 1234
Temperature 80C: 82C
Temperature 150: 150C
Mark 3 Moisture Analyzer
Module SN: M3085010B
Calibration passed 04/10/11 02:05 PM
Signature:
__________________________
Preadjust 80: 79
Adjusted 80: 80
Preadjust 150: 144
Adjusted 150: 150
Temp adjust done
09/07/2012 02:14 PM