Remote Operation (Computer)
SBI functions for remote operation
are useful if the current device/application
status can be read out at any time. This
is achieved using the ESC ars_ command
(see next page). The response from the
device to the ESC ars_ command has the
format: XXX xx zz ff.
41
Device response Meaning
XXX Active application
xx Application status
zz Cover status
ff Function of the u
key
XXX Meaning
MAN Moisture analysis/Setup
menu/Program menu
CBA Cal./adj. scale
CHT Heater adjustment
CMS Measurement
adjustment
THW Hardware test
xx Operating status Action Main display
1 TARE Tare sample pan Weight value
2 WEIGHING Apply sample Weight value
3 WEIGHING Start analysis Weight value
4 TARE/MEASUREMENT START Waiting for stability Dark
5 TARE Waiting for tare to complete Dark
6 START MEASUREMENT Close sample chamber, Weight value
delay time running
7 MEASUREMENT Analysis active Result
8 MEASUREMENT PAUSE Analysis cover open Weight value
9 MEASUREMENT END Analysis ended Result
14 SETUP/PROGRAM Configure parameters Menu display
15 ERROR L, H, scale error 50, 53, 54 L, H, ERR xxx
zz Meaning
O Cover open
C Cover closed
ff Meaning
0 No soft key
1 Soft key 1
2 Soft key 2
3 Soft key 3