Setting the Drying Parameters
Purpose
Adapt the moisture analyzer to the special requirements of products.
Parameters can be configured individually for each program.
Drying Parameters (Overview)
ο Factory settings
√ User-defined setting
Program ο P1
memory to
(
PROGRAM) P20 Program name Blank (max. 16 characters)
Heating program ο Standard drying Temperature 105°C 40 … 220°C for MA150Q,
otherwise 40 … 180°C
Gentle drying Temperature 105°C 40 … 200°C for MA150Q,
otherwise 40 … 180°C
Time 3.0 min. 1,0 … 20 minutes
Standby ο Off
temperature On Temperature 40°C 40 … 100°C
Initial weight ο Off
On Target 5.0 g 0.2 g … 90% max load
Start of analysis With stability
when Enter pressed
ο Without stability
when cover is closed
With stability
when cover is closed
End of analysis ο Fully automatic
Semi-automatic Loss 5 mg 1 … 50 mg
Absolute Time interval 24 s 5 … 300 s
Semi-automatic Loss 1 % 0.1 … 5.0 %
Percentage Time interval 60 s 5 … 300 s
Time Time 15.0 min 0.1 … 99.9 minutes
Manual
Display mode ο Moisture (%L)
Dry weight (%R)
Ratio (%LR)
Weight loss (mg)
Residue (g)
Residue weight (g/kg)
Gram/liter (g/l) Sample 1.0 l 0.0001 … 9.9999 l
Print ο Off
intermediate result On Interval 0.1 min 0,1 … 10.0 minutes
Determination of ο Off
ignition residue On
Headers Line 1 Empty max. 20 characters
Line
2 Empty max. 20 characters
Factory settings No
Yes
24