Features
  Number of Drying Programs:
  20 programs
  Programs are listed according to  
program number.
  Heating Programs
  Two heating programs are available  
for performing moisture analysis of  
a substance:
–  Standard drying
–  Gentle drying
  Standard drying:
  The final temperature needs to be 
entered when using the standard  
program.
 
Standardtrocknung
T [°C]
t [s]
  If a target temperature of over 200°C  
is entered, the moisture analyzer heats 
to the target and then adjusts to 200°C 
in 5 phases, from the 10
th
 to the 30
th
 
minute.
 
Standardtrocknung
T [°C]
10 30 t [min]
200°C
 
  Gentle drying:
  Heat up to the target temperature 
(maximum 200°C) in a selectable time-
frame between 1 and 20 minutes.
25
  Standby Temperature
–  Temperature set to defined value  
when the sample chamber is closed
  Initial Weight
  The target weight (as a guide for  
applying a sample to the analyzer –  
not as a starting condition) can be 
entered by the user. The tolerance  
limits of ± 20% are displayed.
  Start of Analysis 
–  With stability after pressing u key:
  When 
START is shown in the function 
line and you press u to confirm, 
the initial weight is stored at stability 
regardless of whether the cover is  
open or closed.
–  Without stability after the cover  
is closed:
  A prompt to close the cover  
(a graphic symbol) is displayed once  
the initial weight condition is met.
  The initial weight is stored without  
stability as soon as the sample chamber 
is closed.
  The analysis starts as soon as a 2-second 
delay has elapsed after the cover is 
closed or the sample chamber is opened 
and then closed (applies to both start 
criteria).
–  With stability after cover is closed:
  A prompt to close the cover (a graphic 
symbol) is displayed once the initial 
weight condition is met.
  The initial weight after stability is  
stored as soon as the sample chamber  
is closed. There is no additional  
delay time. 
–  Interruption of analysis: 
  You can open and close the cover  
at any time without interrupting the 
measurement to post-treat samples.
  The device status line shows  
MEASUREMENT PAUSE/MEASURE-
MENT when you open/close the cover.  
The drying and heating algorithms stop 
when the cover is open and continue 
once it is closed.
  End of Analysis with Shutoff  
Parameters
–  Fully automatic
–  Semi-automatic, absolute
–  Semi-automatic, percentage 
–  Timer mode
–  Manual
  Fully automatic:
  The analysis ends as soon as the weight 
loss per 24 s is below an automatically 
detected threshold.
 
Weight
Moisture
Automatic shutoff
Time
  Semi-automatic, absolute:
  The analysis ends as soon as the weight 
loss per 24 s is lower than a user-
defined threshold (set in milligrams). 
  Semi-automatic, percentage:
  The analysis ends as soon as the weight 
loss per 24 s is lower than a user-
defined threshold (set in percent). The 
percentage is entered by the user.
  Timer mode:
  The analysis ends as soon as the  
specified time has elapsed.
  Manual:
  Shut-off with the u key.