Display Mode
The following units can be selected for
displaying analysis results:
Moisture %L
Dry weight %R
Ratio %LR
Weight loss mg
Residual weight g
Residual weight g/kg
Gram/liter g/l
Print Intermediate Results
When set to “Off”, intermediate
results can only be printed by pressing
the r key.
Intermediate results can be printed out
automatically after a configurable time
interval.
Determination of Ignition Residue
A two-part measurement can be
performed using this setting. Example:
Moisture analysis and incineration
of sludge. Moisture content is analyzed
in the first stage. The sample is then
reduced to ashes in the furnace and
in the second stage the ash is calculated
as a percentage of the initial weight.
CONTINUE is displayed instead of END
after the first stage.
The moisture analyzer must not be used
for other purposes between the first and
second stages.
Headers
Text can be entered for the drying
report (2 lines each with max.
20 characters). The relevant line is
not printed if no text is entered.
Factory Setting
The drying programs can be reset to
the factory settings.
Printout of the Analysis Parameters
§ When the desired program is displayed:
Press the r key
> Printout (example on right). Lines with
more than 20 characters are truncated.
Additional Functions
The following functions are supported
in the program memory:
Display programs
Change settings
Save program changes
-------------------- 
14.07.2006     12:40 
Model MA150Q-000230V 
Ser. no.    18701636 
Vers. no.   01-50-01 
ID 
-------------------- 
      PROGRAM 
-------------------- 
 P1 
  Program name 
   Progr. name: 
               MEYER 
  Heating program 
   Standard drying 
    Temperature: 
                105 °C
  Standby temperature 
                  Off 
  Initial weight 
                  Off 
  Start of measuremen 
Without stability,  
          after close 
  End of measurement 
      Fully automatic 
  Result display 
Moisture         (%L) 
  Print intermediate 
                 res. 
                  Off 
  Determination of  
     ignition residue 
                  Off 
  Headers 
   Line 1 
    Line 1: 
   Line 2 
    Line 2: 
--------------------
26