Data Output Format
The contents of the measured value
line and the unit are output with data
ID codes.
Example
N + 3.4253 g
20 characters per line are output.
Output format with 20 characters + CR LF
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
I I I I I I + * D D D D D D D D * U U U CR LF
* * * * * – . . . . . . . . * * *
* * * * * * * * *
I: ID code character U: Unit symbol
*: Space CR: Carriage return
D: Display character LF: Line feed
Error Code
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * E R R * # # # * * * * CR LF
*: Space # # #: Error code number
H: Overload
L: Underload
Digital outputs (status display, programmable logic controller)
The respective status of the moisture analysis is output to digital output pins 16–19
(see pin assignment chart) as per the following table.
Status Pin16 Pin17 Pin18 Pin19 Description
No moisture analysis 0 0 0 0 No drying program
(OFF, STANDBY, SETUP, PROG,
ERR xxx)
TARE inactivity 0 0 1 0 Wait for u key “TAR”
TARE waiting for 0 0 1 1 key u, “TAR” confirmed,
tare to complete Waiting for stability
WEIGH IN 0 1 0 0 Measured value display
Apply sample
WEIGH IN 0 1 0 1 Measured value display
Start measurement
MEASUREMENT PAUSE
MEASUREMENT started 0 1 1 1 Measured value display
MEASUREMENT 0 1 1 0 Measured value display
Reading display
MEASUREMENT END 1 0 0 0 Measured value display
MEASUREMENT CANCEL
1 1 1 0 Measured value display
40