Service
  Regular servicing by a Sartorius  
technician will extend the service life  
of your moisture analyzer and ensure  
its continued weighing accuracy.  
Sartorius can offer you service contracts 
with your choice of regular maintenance 
intervals ranging from 1 month to  
2 years.
  Maintenance intervals depend on  
operating conditions and individual  
tolerance requirements.
  Repairs
  Repair work must only be carried out  
by trained service technicians.  
Any attempt by unqualified personnel 
to carry out repair work may result  
in hazards for users.
  Cleaning
  Depending on the type of sample,  
cooling vapors may condense on the 
cover and result in discoloring of  
the cover. This is not detrimental to  
the operation of the device.
! Make sure that no dust or liquid pene-
trates the moisture analyzer housing
! Do not use any aggressive cleaning 
agents (solvents, abrasive cleaning 
agents, etc.); clean the moisture  
analyzer with a cloth moistened with  
a mild detergent (soap) only
§  Isolate from supply voltage: Unplug 
the power cord from the wall outlet 
(mains supply). Unplug any connected 
data cables from the analyzer
$  The pan draft shield and the pan 
support can be removed for cleaning
§  Carefully remove any sample residue/
spilled powder using a brush  
or hand-held vacuum cleaner
§  Wipe down analyzer with a soft, 
dry cloth after cleaning
  Cleaning the Heating Unit and  
Temperature Sensor
§  Open the hood
! Danger: The terminals of the heating 
unit are under live current
§  Isolate from Supply Voltage: 
Unplug Power Cord. 
  Disconnect data cables from  
moisture analyzer
§  Carefully remove any residues from 
the temperature sensor
§  Use a soft-bristled brush or a commer-
cially available cleaning agent for  
glassware to clean the ceramic heating 
element.
§  Use a commercially available glass 
cleaner to clean the protective glass  
on the quartz-rod radiator.
48
Care and Maintenance