EasyManua.ls Logo

Sartorius MA150 - Overview and Specifications; Detailed Technical Specifications

Sartorius MA150
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Dryer functions:
Model MA150C MA150Q
Heating element Ceramic IR heater Quartz-rod IR heater
Temperature range 40 to 180 °C 40 to 220 °C
Temperature increments Adjustable in 1 °C increments
Temperature adjustment With YTM03MA/YTM05MA temperature adjustment set
Weighing functions:
Weighing capacity 150 g
Readability 1 mg, 0.01% moisture content
Repeatability, typical (%) Sample weight 1 to 5 g: 0.2%
Sample weight 5 g and over: 0.05%
External calibration weight 100 g (E2)
(with an accuracy of at least…)
Sample pan dimensions d 90 mm
Drying parameters:
Heating programs Standard; gentle drying
Drying time 0.1 to 99.9 minutes
Number of stored programs 20
Shutoff criteria Fully-automatic, semi-automatic (absolute, percentage),
timer mode (1 + 99.9 min.), manual
Display of analysis results Moisture, dry weight, ratio, weight loss, residue (g or g/kg), g/l
(with specification of sample quantity)
Minimum sample weight 0.100 g
Analyzer (hardware):
Dimensions (W + D + H) 213 + 320 + 181 mm
Net weight, approx. 5.1 kg
Voltage 230 V or 115 V (depending on model variant), –15% to +10%
Frequency 48 60 Hz
Fuses 2 (neutral conductor/phase), 6.3 AT, 5 + 20 mm (internal)
Operating temperature range +10 … +30 °C
Power consumption 700 VA max.
Built-in interface RS-232C
Format: 7 bit ASCII, 1 start bit, 1 or 2 stop bits
Parity: odd or even
Transmission rates: 150 to 19,200 baud
Handshake: Software or hardware
Digital input: 1, configurable function
Digital outputs: 4, moisture analysis status
50