Display Cause Remedy
What should I do if The selected temperature is too high Reduce drying temperature
and the sample oxidizes Put a glass-fiber filter on top of
Sample boils or scorches and the splashes the sample
continually change the sample weight Reduce the sample quantity or
distribute more evenly
Select a semi-automatic shutoff
parameter or select timer mode
Analysis time too long Increase temperature
Reduce sample quantity
Dry 2 to 3 minutes with empty
disposable sample pan (pre-heat)
Sample loses weight prior to start Remove the sample pan and apply
of analysis sample outside the sample chamber
Sample is liquid or pasty Use a glass fiber filter
Sample has only a low moisture content Increase sample quantity
Insufficient heat output Clean temperature sensor
Place of installation unstable (vibrations, etc.) Change the place of installation
Please contact the Sartorius Service Center if any other problems occur.