34
Moisture Analysis
Operating Instructions Moisture Analyzer MA37
Moisture in %M
Standard drying 105 °C
Standby temperature -
Initial weight -
End Fully automatic
Target value -
No automatic printout
y The menu is displayed.
t Select the M button in the menu.
Customize parameters
Heating program
End
Initial weight
Result display
Standard drying
105 °C
Fully automatic
Off
Moisture in %M
Target value
Off
y The current method parameters are displayed.
Customize parameters
Heating program
End
Initial weight
Result display
Standard drying
105 °C
Fully automatic
Off
Moisture in %M
Target value
Off
t Select the parameter you wish to change, such as “Heating program.”
Heating program
Type Temperature [°C]
110
85
125
Gentle drying
Standard drying
t Set the parameters to suit your requirements.
Select t if necessary and enter the desired values (see “Entering Numbers on
the Number Pad” in Chapter “Operating the Moisture Analyzer”).
t Select l to confirm.