MA Performance Test
Operating Instructions Moisture Analyzer MA37
MA performance test
Device is tested with reference sample.
Temperature: 70 °C
Duration: 05:00 min
00:13 min
y The display shows the progress of the test and the remaining time.
Performance Test Measurement Not Successfully
Completed
MA performance test
Measurement canceled
y The message “Measurement canceled” appears.
t Select l to confirm.
y The display switches back to the menu.
If a fault has been detected:
MA performance test
Device is not OK.
5 MApi << 10 MApi
Temperature 70 °C
Please contact Sartorius service
a
y The message “Device is not OK” appears.
t Select m to confirm.
Contact Sartorius service if the message “Device is not OK” appears.
Performance Test Successfully Completed
al
MA performance test
Device is OK.
Result 10 MApi
Temperature 70°C
3
y If the test is successful, the message “Device is OK” first appears. Then the
analyzer indicates that the sample can be removed.
Remove old sample
t Select l to confirm.
y You can remove the sample and return to the menu.