Operating Instructions Moisture Analyzer MA37
+ 2.500 g
Place pan, close hood for taring
t Place a sample pan on the analyzer.
t Close the hood.
y The moisture analyzer tares automatically.
Open hood
0.000 g
t Open the hood.
t Open the packaging of the ReproEasy pad (accessory) and remove the pad.
t Remove the protective wrapping.
t Affix the pad to the center of the sample pan, sticky side down.
Fill in reference sample and
close hood to start
+ 2.513 g
t Place the ReproEasy pad reference sample (accessory) on the sample pan.
Close hood to start
+ 6.721 g
t Close the hood.
y The performance test starts automatically.
During the performance test, the process status light on the analyzer flashes,
if the light has not been switched off in the menu.