46
ISO/GLP-compliant Printout
Operating Instructions Moisture Analyzer MA37
ISO/GLP-compliant Printout
Features
You can have the device information, device ID, and the current date printed
before (GLP header) and after (GLP footer) the values from the weighing series.
The following data is printed:
GLP header:
Date, time at the start of a weighing series
– Manufacturer
– Model
Model serial number
Software version number of the device (BAC)
Software version number of the display (APC)
Software version number of the heating module (DOC)
Device ID with max. 14 characters (if activated under system settings)
GLP footer:
Field for signature
Configuration
To print out the ISO/GLP report, define the following system settings
(see Chapter “System Settings”, section “Printout”):
t Activate ISO/GLP-compliant printouts:
In the Printout menu under Printout and Report Format, activate the
With GLP and method parameters option.
Operation
t Print out the report after moisture analysis: Select a.