EasyManua.ls Logo

Sartorius MA37 - Specifications

Sartorius MA37
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
62
Specifications
Operating Instructions Moisture Analyzer MA37
Specifications
Dryer Functions MA37-1
Temperature range and settings 40°C–160°C, in increments of 1°C; stand-by temperature selectable from 40–100°C
Sample heating
Infrared heating using a metal tube heater
Heating programs Standard drying, gentle drying
Switch-off criterion Optional:
– Fully automatic
– Semi-automatic mg (1–50 mg/5–300 s)
– Semi-automatic % (0.1–5.0%/5–300 s)
– Time (2:00–99.59 min.)
– Manual
Displayed result Moisture in %M and g; dry weight in %S and g; ATRO in %M/S
Net weight, approx. 6.3 kg
Weighing Functions
Max. weighing capacity 70 g
Repeatability, average For initial sample weight approx. > 1 g: ± 0.2%
For initial sample weight approx. > 5 g: ± 0.05%
Readability 1 mg; 0.01%
Typical sample quantity 5 to 15 g
External calibration weight 50 g (E2)
Disposable sample pan dimensions d 90 mm
Interface Mini USB: Automatic detection of Sartorius printers YDP30 and YDP40,
direct data transfer to Microsoft
®
Windows programs without additional software,
programmable data output interval
Power Supply
Input supply voltage
100–120 V AC and 210–240 V AC (automatic voltage detection), ±10%, 50/60 Hz
Protection class and standards
Protection class I as per EN 61010-1/IEC 61010-1; up to 3000 m above sea level; IP20 as per
EN 60529/IEC 60529
Power supply Using country-specific power cable from Sartorius only
Power consumption Max. 640 W (typically)
Ambient Conditions
The specifications are valid for the following ambient conditions:
Environment For indoor use only
Ambient temperature +10°C to +30°C
Operating temperature Operation guaranteed between +5°C and +40°C
Storage and shipping –10°C to +60°C
Height Up to 3000 m above sea level
Relative humidity 15% to 80% for temperatures up to 31°C non-condensing,
decreasing linearly up to 50% relative humidity at 40°C and 20% at 50°C
Safety of Electrical Equipment
As per EN 61010-1/IEC 61010-1
Safety requirements for electrical equipment for measurement, control, and laboratory use —
Part 1: General requirements
Electromagnetic Compatibility
As per EN 61326-1/IEC 61326-1
Electrical equipment for measurement, control, and laboratory use — EMC requirements —
Part 1: General requirements
Defined immunity to interference: Suitable for use in industrial areas
Limitation of emissions: Class B [suitable for use in residential areas and areas that are connected to a low voltage
network that (also) supplies residential buildings]. Therefore, the
analyzer can be used in both areas.