Error Messages
Certain events cause a message to appear on the display:
1. Some error messages are displayed for approx. 3 seconds
(e.g., CAL-Extern: calibration canceled).
2. Some error messages are displayed until they are acknowledged
by pressing a key (e.g., “Incorrect access code” in “Service access”).
3. Information messages are displayed until they are acknowledged
by pressing a key.
Deactivated Keys
To prevent operating errors, only relevant functions/keys are displayed depending
on the situation. This prevents most operating errors from occurring. The following
buttons are only available in specific circumstances:
A
Only if there is a measured value in the zero setting range before
external calibration/adjustment.
D
Only if there is a measured value larger or smaller than “zero.”
Error Messages during Operation
“Value is too small.” When an entered value is too low for the parameter.
“Value is too large.” When an entered value is too high for the parameter.
Error 46 When the max. heating temperature has been briefly
exceeded.
Error 401 to 407 Faulty heating element. Contact the Sartorius Service
Center.