Calibration and Adjustment
Background During calibration, a calibration weight is used to determine how much the
displayed value deviates from the actual measurement value. This deviation is
compared against a preset target value. The subsequent adjustment corrects
this deviation or reduces the permissible error limits.
Calibration and adjustment are combined in one process in the moisture analyzer.
After each calibration the analyzer is automatically adjusted.
When and How Often The moisture analyzer must be calibrated and adjusted, for example:
– When the ambient conditions have changed (temperature, humidity, or air
pressure)
– When the moisture analyzer is set up in a different location or moved in its
current location.
An external calibration weight is required for calibration and adjustment (100 g).
Please note the tolerance of the calibration weight being used.
Calibrating the Moisture Analyzer
t Make sure that the pan support is empty.
t Select the menu key on the main screen.
Moisture in %M
Standard drying 105 °C
Standby temperature -
Initial weight -
End Fully automatic
Target value -
No automatic printout
y The menu is displayed.
t Select the i (Setup) key in the menu.