y The Settings menu appears.
t Select Calibration/Adjustment.
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
t Select Weighing system adjustment.
t Select CAL-Extern to calibrate the analyzer using an external calibration
weight.
y The Calibration/Adjustment window appears.
t Close the hood and select l to confirm.
t The CAL-Extern window appears.
t If necessary, please zero the integrated balance.
y A message appears on the display, prompting you to place the calibration
weight (50 g) on the analyzer.
t Place the calibration weight on the pan support.
y Calibration starts automatically.
When calibration is complete, a report appears on the display. The report gives
the deviation detected during calibration. You are asked whether the weighing
system should be adjusted. After adjustment the report shows the result of the
adjustment.
t Close the report: Select <.
y The moisture analyzer is now calibrated. If you have also carried out an
adjustment after calibration, the moisture analyzer is also now adjusted.
CAL-Extern
START
Please set balance to zero
G - 0.401 g
CAL-Extern
Please put on required weight
Close hood
Stat - 50,000 g