Customize parameters
Initial weight
Off
Result display
Target value
Standby temperature
Moisture in %M
Off
Off
t Scroll through the list if necessary and change further parameters.
t When you are finished making changes, select o.
Save changes
Temperature: 110 °C
Initial weight: On
Target value: On
NOYES
y The changed method parameters are displayed.
t Select the desired option:
– Make further changes: Select <. This takes you back to the list of
parameters.
– Reject changes: Select NO.
– Save changes: Select YES.
y The menu is displayed.
t To return to the main screen, select k.
Performing the Moisture Analysis
Moisture in %M
Start new measurement
Standard
Fully automatic
105 °C
Wednesday, 2014-01-08 16:50
When all parameters have been set to suit your requirements, you can start the
moisture analysis.
Observe the following instructions in order to ensure successful measurement:
– Wear gloves for each measurement, so that the liquid in the sample is not
influenced by handling.
– Remove the sample directly from the process or keep it in an airtight and
watertight container until measurement.
– Ensure that the sample is representative for the product and is homogeneous
(shake, stir, or mix before measurement).
– If using a filter or filters, store the filters in airtight and watertight packaging.
t Select START on the main screen or open the hood.
If you selected START on the main screen:
Open hood
0.000 g