8
Startup
Operating Instructions Moisture Analyzer MA37
Startup
Unpacking and Equipment Supplied
t Open the packaging, making sure to remove all parts carefully.
t After unpacking the analyzer, check it immediately for any external damage.
If you detect any damage, proceed as directed in “Care and Maintenance.”
Save the box and all parts of the packaging for any future transport. During
shipment, please do not leave cables plugged in!
The following parts are included in the equipment supplied:
Moisture analyzer
Country-specific power cable
Pan support
Protective cover for the control panel
80 disposable aluminum sample pans
80 glass-fiber filters
Operating instructions
Choosing a Location
t Select the proper setup location:
An optimal location guarantees accuracy and reliability. Ensure that the following
ambient conditions are met:
Set up the analyzer on a stable, even surface that is not exposed to vibrations
(e.g., weighing stone).
Maintain free access to the analyzer at all times.
Make sure to choose a place where excessive heat cannot build up.
Leave enough space between it and materials that are affected by heat.
Choose a location that is not subject to the following negative influences:
Heat (heater or direct sunlight)
Drafts from open windows, AC systems, and doors
Extreme vibrations during measurement
Heavy traffic areas (personnel)
Extreme humidity
Electromagnetic fields
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it is
brought into a substantially warmer area. You should therefore let a device that has
been disconnected from its power source acclimatize for approximately 2 hours
before reconnecting it to the power.