Calibration and Adjustment
Operating Instructions Moisture Analyzer MA37
Viewing the Calibration Report
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
t To view the most recent calibration reports: Go to the Calibration/Adjustment
menu and select the Calibration report submenu.
Here you can select the most recent reports for the MA performance tests
or weighing system adjustments and then access the most recent calibration
reports, which are sorted by date.
Up to 99 entries per day are saved. Calibration reports older than 30 days are
deleted.
y Menu Calibration report.
Example of a Calibration Report:
--------------------
2014-07-29     12:02
    Sartorius
Mod.          MA37-1
SerNo.    0031710609
BAC:        00-53-01
APC:        01-80-01
DOC:        07-40-03
--------------------
2014-07-29      12:02
External calibration
Start: manually
Set   +   50.000 g
Dev   -    0.002 g
External adjustment
Dev        0.000 g
Temp  +     23.7 °C
--------------------
Name:
--------------------
Calibration Report